PHOTOELECTRIC ANALYZER OF CONTAMINATION IN LIQUID
AZG-975



Contamination analyzer AZG-975 is intended for the automatic monitoring of mechanical impurities content in hydraulic, oil and fuel lines of aircrafts and machinery. Analysis is performed in samples taken. AZG-975 maintains information exchange with PC and remote control. Results can be represented according to GOST 17216-71, ISO 4406, NAS 1638.


Azg-975



SPECIFICATION

Liquids controllable - bensene, cerosene, fuels, oils etc. with viscosity at T = + 20o Ñ, not more, cSt 20
Volume of liquid sample, not more, ml 130
Size ranges of contamination particles (diameter), 10-6m 5... 10
10... 25
25... 50
50... 100
100...200
more than 200
Maximium particle concentration in sample, not less, cm-3 1500
Particle count error in range 100-200, not more, % 30
Consumed power, not more, W 7
Dimensions, mm:
- remote sensor with glass
- analyzer

120x230
210x78x210

It is recommended to use AZG-975 with auxiliary equipment KVANT-904, intended for laboratory dish washing and for vacuum providing.